This kit includes:
- 1x Carbon dioxide (CO2) Sensor
- 1x 12V power source
- 1x Octal I/O adapter
- 1x Easy wire connector 6 ways
The assembly is pre-cabled with wires and adapters.
Measurement performance 1)
Relative humidity
Measurement range 0 … 100 %RH
Accuracy 2) 3)
At 0 … +40 °C (+32 … +104 °F) ±1.5 %RH (0 … 90 %RH)
±2.5 %RH (90 … 100 %RH)
At −40 … 0 °C and +40 … +80 °C
(−40 … +32 °F and +104 … +176 °F)
±3.0 %RH (0 … 90 %RH)
±4.0 %RH (90 … 100 %RH)
Factory calibration uncertainty at
+20 °C (+68 °F)
±1.1 %RH (0 … 90 %RH)
±1.8 %RH (90 … 100 %RH)
Humidity sensor types HUMICAPâ 180R
HUMICAPâ 180V
Stability ±2 %RH over 2 years
Stability in typical HVAC applications ±0.5 %RH per year
Temperature
Measurement range −40 … +80 °C (−40 … +176 °F)
Accuracy over temperature range:
At +15 … +25 °C (+59 … +77 °F) ±0.1 °C (±0.18 °F)
At 0 … +15 °C and +25 … +40 °C
(+32 … +59 °F and +77 … +104 °F)
±0.15 °C (±0.27 °F)
At −40 … +0 °C and +40 … +80 °C
(−40 … +32 °F and +104 … +176 °F)
±0.4 °C (±0.72 °F)
Temperature sensor Pt1000 RTD Class F0.1 IEC 60751
Other output parameters (optional)
Dew point/frost point, wet bulb temperature, enthalpy, absolute humidity,
mixing ratio, vapor pressure, saturation vapor pressure
1) RH specifications excluding TMT120 and TMT130, which are T-only models.
2) Including non-linearity, hysteresis, and repeatability.
3) With HUMICAPâ 180V sensor, accuracy is specified only in operating temperature −20 ... +80 °C
(−4 ... +176 °F).